请教:金宝200S血透机故障处理!!!!!!
请教:金宝200S血透机开机后面板上时间显示从0.15开始倒计时,当倒计时到案.09时立刻又回到0.15开始倒计时至0.09,如此重复上述,显示屏上显示故障代码 ERROR! RESTART OR CALL SERVICE ADC 4.01 009 可能是HPG压力传感器有问题 ADC 4.01 0xx = More then 20% over range on channel xxChannel xx Explanation
0 MAIN PC BOARD TEMPERATURE
1 MOTOR CURRENT
2 VALVE CURRENT
3 CONDUCTIVITY CONTROL ACETATE/TOTAL
4 DEGASSING PRESSURE
5 HEATING VESSEL TEMPERATURE
6 CONDUCTIVITY CONTROL BCM OPTION
7 TEMPERATURE BCM CONDUCTIVITY TRANSDUCER
8 UF TRANSDUCER COIL CURRENT
(POSITIVE SYNC)
9 HIGH PRESSURE GUARD (HPG)
10 DIALYSIS PRESSURE (PD)
11 AUX INPUT
12 AUX INPUT
13 AUX INPUT
14 OFFSET VOLTAGE AD-SYSTEM
15 REFERENCE VOLTAGE AD-SYSTEM
所以应该是HPG的问题, 本帖最后由 sdu2006 于 2011-11-21 09:09 编辑
ADC 4.01 0xx = More then 20% over range on channel xx
Channel xx Explanation
0 MAIN PC BOARD TEMPERATURE
1 MOTOR CURRENT
2 VALVE CURRENT
3 CONDUCTIVITY CONTROL ACETATE/TOTAL
4 DEGASSING PRESSURE
5 HEATING VESSEL TEMPERATURE
6 CONDUCTIVITY CONTROL BCM OPTION
7 TEMPERATURE BCM CONDUCTIVITY TRANSDUCER
8 UF TRANSDUCER COIL CURRENT
(POSITIVE SYNC)
9 HIGH PRESSURE GUARD (HPG)
10 DIALYSIS PRESSURE (PD)
11 AUX INPUT
12 AUX INPUT
13 AUX INPUT
HPG传感器故障
14 OFFSET VOLTAGE AD-SYSTEM
15 REFERENCE VOLTAGE AD-SYSTEM 学习了,谢谢分享! shuaism 发表于 2011-3-23 22:41
ADC 4.01 0xx = More then 20% over range on channel xx
Channel xx ...
有金宝AK200S的BM主板出售吗?有的请联系QQ929909703谢谢!