lkyl 发表于 2011-3-18 21:50:57

请教:金宝200S血透机故障处理!!!!!!

请教:金宝200S血透机开机后面板上时间显示从0.15开始倒计时,当倒计时到案.09时立刻又回到0.15开始倒计时至0.09,如此重复上述,显示屏上显示故障代码 ERROR! RESTART OR CALL SERVICE ADC 4.01 009

金宝配件 发表于 2011-3-23 22:20:25

可能是HPG压力传感器有问题

shuaism 发表于 2011-3-23 22:41:19

ADC 4.01 0xx             =   More then 20% over range on channel xx
                                     Channel xx         Explanation
                                   0         MAIN PC BOARD TEMPERATURE
                                               1         MOTOR CURRENT
                                               2         VALVE CURRENT
                                             3         CONDUCTIVITY CONTROL ACETATE/TOTAL
                                               4         DEGASSING PRESSURE
                                                5         HEATING VESSEL TEMPERATURE
                                               6         CONDUCTIVITY CONTROL BCM OPTION
                                               7         TEMPERATURE BCM CONDUCTIVITY TRANSDUCER
                                            8         UF TRANSDUCER COIL CURRENT
                                                                (POSITIVE SYNC)
                                               9         HIGH PRESSURE GUARD (HPG)
                                              10      DIALYSIS PRESSURE (PD)
                                                 11      AUX INPUT
                                                12      AUX INPUT
                                                13      AUX INPUT
                                             14      OFFSET VOLTAGE AD-SYSTEM
                                                 15      REFERENCE VOLTAGE AD-SYSTEM
所以应该是HPG的问题,

sdu2006 发表于 2011-11-21 09:07:00

本帖最后由 sdu2006 于 2011-11-21 09:09 编辑

ADC 4.01 0xx             =   More then 20% over range on channel xx
                                     Channel xx         Explanation
                                   0         MAIN PC BOARD TEMPERATURE
                                               1         MOTOR CURRENT
                                               2         VALVE CURRENT
                                             3         CONDUCTIVITY CONTROL ACETATE/TOTAL
                                               4         DEGASSING PRESSURE
                                                5         HEATING VESSEL TEMPERATURE
                                               6         CONDUCTIVITY CONTROL BCM OPTION
                                               7         TEMPERATURE BCM CONDUCTIVITY TRANSDUCER
                                            8         UF TRANSDUCER COIL CURRENT
                                                                (POSITIVE SYNC)
                                               9         HIGH PRESSURE GUARD (HPG)
                                              10      DIALYSIS PRESSURE (PD)
                                                 11      AUX INPUT
                                                12      AUX INPUT
                                                13      AUX INPUT


HPG传感器故障
                                             14      OFFSET VOLTAGE AD-SYSTEM
                                                 15      REFERENCE VOLTAGE AD-SYSTEM

yj7107 发表于 2012-2-1 23:57:56

学习了,谢谢分享!

chenxi4576698 发表于 2018-3-19 20:43:43

shuaism 发表于 2011-3-23 22:41
ADC 4.01 0xx             =   More then 20% over range on channel xx
                                     Channel xx          ...

有金宝AK200S的BM主板出售吗?有的请联系QQ929909703谢谢!
页: [1]