请教金宝血透AK95,9V电池问题
遇到一台金宝AK95血透机器,错误代码C COFF 088 095,代码提示9V电池故障,但是更换新电池后,此错误依旧出现。请教各位大侠,遇到此类情况,如何解决,真心感谢! C COFF 088 095 UNSTABLE CONDUCTIVITY (HIGH TEST)
Unconditional error, blood pump available
General description of conditions for occurring:
- No conductivity value sent from CPU C to CPU P due to unstable signal
Technical description of conditions for occurring:
- PRI(PI_CondNoAchHighErrCnt) > 0
- Unstable cond. value in conductivity cell B, FRI(FI_MM90CONDB)
Some possible explanations why the error has occurred:
- Air in fluid path
- Broken CPU C conductivity cell (cell B)
- Broken MM I/O board
- Broken control driver board 回复 2# 季如宁
错误代码 C COFB 088 096电池故障,开机后充不了电,9V电池迅速降到7.5以下! 电池座有没氧化,导致接触不好。如连接没问题,换保护驱动板试试 088 095 不是电池,是电导块问题。088 096 才是9 V电池故障 销售金宝血透机电导度传感器,价格仅为原厂1/3, 手机:13961395802