hhxvgg 发表于 2011-6-23 16:21:48

请教金宝血透AK95,9V电池问题

遇到一台金宝AK95血透机器,错误代码C COFF 088 095,代码提示9V电池故障,但是更换新电池后,此错误依旧出现。
请教各位大侠,遇到此类情况,如何解决,真心感谢!

季如宁 发表于 2011-6-23 21:42:16

C COFF 088 095        UNSTABLE CONDUCTIVITY (HIGH TEST)
        Unconditional error, blood pump available
       
        General description of conditions for occurring:
        - No conductivity value sent from CPU C to CPU P due to unstable signal
       
        Technical description of conditions for occurring:
        - PRI(PI_CondNoAchHighErrCnt) > 0
        - Unstable cond. value in conductivity cell B, FRI(FI_MM90CONDB)
       
        Some possible explanations why the error has occurred:
        - Air in fluid path
        - Broken CPU C conductivity cell (cell B)
        - Broken MM I/O board
        - Broken control driver board

hhxvgg 发表于 2011-6-24 08:57:12

回复 2# 季如宁


    错误代码 C COFB 088 096电池故障,开机后充不了电,9V电池迅速降到7.5以下!

金宝配件 发表于 2011-6-25 09:54:58

电池座有没氧化,导致接触不好。如连接没问题,换保护驱动板试试

星空牧羊 发表于 2013-1-27 01:46:53

088 095 不是电池,是电导块问题。088 096 才是9 V电池故障

fsy6612 发表于 2013-5-5 22:54:55

销售金宝血透机电导度传感器,价格仅为原厂1/3, 手机:13961395802
页: [1]