求助:GAMBRO AK95S CCFM 020 015
如题:GAMBRO AK95S CCFM 020 015, 期间不断 FLUID PATH OBSTRUCTION, 请问如何解决??? 反复除钙,不行的话可能要换控制驱动板,我们换过 坏消息, UF CELL 需要更换,价钱还蛮给力。 按RESET或许可以解决 Some possible explanations why the error has occurred:- Air is flowing through flow channel 1. There is too much air in the inlet water
- Valves: RIVA, ZEVA, diva or TAVA are leaking
- Flow cell: the flow cell is dirty.
- Flow cell: the flow cell is broken
- Flow restrictor: the restrictor is dirty
- Flow pump: the pump is not working properly
- Degassing chamber: the o-ring at the top of the chamber is not in its correct
position. fluid is leaking.
- Degassing pressure transducer: not working properly.
不需要翻译了吧?:lol 那就要清洗了,多洗洗就好了,实在不行就要换超滤单元了 能翻译一下,最好。。