求助:GAMBRO AK95S C CC 008 104
GAMBRO AK95S 开机报扳手故障,显示CCC 008 104 CCC 008 105 CCC 008 106 三个故障!请高手指教 这个不太懂 高手呢,出来帮下忙了! ERROR READING PARAMETERS FOR COND A TEMP.从电导度传感器A上的TEMP温度传感器读取参数时发生错误。
Unconditional error, blood pump not available
General description of conditions for occurring:
- An error occurred while reading transducer parameters during init (start-up).
格式化传感器读取数据时发生错误。
Some possible explanations why the error has occurred:
- No connection between transducer and MM95 I/O board
传感器板没有接到 MM95 I/O 板上。
- Incorrect connection between transducer and MM95 I/O board
传感器板与MM95 I/O板之间的连接有问题。
- No connection between MM95 I/O and control CPU board
MM95 I/O 板与控制CPU 板之间没有连接。
- Incorrect transducer ID. Update transducer ID (I2C scan).
不正确的传感器地址(ID),重新地址扫描。(I2C)
- EEPROM on transducer board is broken传感器板上EEPROM的有故障。
- Serial bus error on control CPU board, or I/O SIMM problem (P310)
控制CPU板上的总线有问题,或I/O SIMM 有问题(P310) 谢谢楼上的,故障处理了!
是A板的地址丢失,重新更换A电路板故障排除! 谁有95的故障代码可以分享一下吗?中文最好,英文也行