richear 发表于 2012-4-5 08:44:26

西门子Sensation64不能完成checkup

————
Function “Filament adaptation” aborted by error。
————

richear 发表于 2012-4-5 09:07:47

3049 CT_XRS         control error XRS
3041 CT_MCU      control error MCU

素茶一杯 发表于 2012-4-5 16:03:28

首先考虑做灯丝校正

22656758 发表于 2012-4-5 19:45:45

控制错误 查下CAN通讯   

richear 发表于 2012-4-6 18:29:19

checkup每次做到120KV495mA这一步就不能往下了。

素茶一杯 发表于 2012-4-11 10:08:51

低KV低mA扫描试试,可以不?会不会和球管有关?

number3 发表于 2012-4-12 00:21:14

Message:
E
CT_XRS_49
CT_XRS_E_50_31__P1P2

Text:
Control error XRS (%1)

Explanation:
UI_LOAD_BAD error is detected.

Possible cause:
- Tube defective
- Generator adaptation incorrect
- Inverter defective
- HV_TANK defective
- Cable connection not OK

Action:
1. Check connection D400 X8 and X9.
2. Perform Generator Adaptation.
3. Perform HVC01 Test with "Test Plugs".
4. If OK, getter/replace tube.
5. If not OK, Use HVC03 Single Pulse Test.
6. If the Shape is OK, replace the HV_TANK.
7. If the shape is not OK, replace the Electronic box.

Information:
FWtext: UI load

P1:n.a.

P2:n.a.

richear 发表于 2012-4-12 12:16:21

球管当掉了,呵呵。
救不过来了。
校正也做不了了,扫描也玩不成了。

richear 发表于 2012-4-13 14:12:01

本帖最后由 richear 于 2012-4-17 11:02 编辑

XRS
D400当了。
球管也当了。

re_jodz 发表于 2012-4-17 11:23:26

球管曝了多少了?