德尔格呼吸机savina300 报警51.0003
报警51.0003 求助各位大神 什么原因 换块主板试试看 这个问题基本是修不好 51.0000 GoldCap capacity is too low.− Replace Control PCB.51.0001 GoldCap test device is faulty.− Replace Control PCB.
51.0002 GoldCap power supply is too high.− Replace Control PCB.
− Carry out SW update ≤1.10.
51.0003 GoldCap power supply is too low.− Replace Control PCB.
51.0004 GoldCap resistance is too high.− Replace Control PCB. 太感谢楼上的师傅们 辛苦了谢谢 辛苦了谢谢