费森4008B故障代码Rinse failure F14
以前出现过,没管它自己好了,现在又出现了,没碰过这机器,不知道怎么下手。 F14 pos. PressureZero point change after closing of solenoid valve V43 (20 sec).
Standard: P-Dial. not equal to –125 to 55 mmHg.
HDF option: P-Dial. not equal to –125 to 60 mmHg.
– The solenoid valve V24 is not closed.
The balancing chamber is switched to passage during this test sequence. V24, V26 and
V43 are closed; V24B is open. 回复 2# xxdxuliang
楼主帮助翻译一下啊。 F14 Shortly before the end of the mandatory rinse in Dis I to V.
The hydraulics test has not been completed correctly, possibly caused
by flow problems. xxdxuliang
手册我也有啊,也知道是流量问题,看来还得拆开来看看了。 问题不是很大,维修模式看看FLOW是否正常? 什么时候的机器?有没有diasafe?
压力传感器、超滤泵、几个阀都看一下 进液过滤器是否脏了堵了 努力学习中!